What Is the Analytical Working Range of an Instrument?
Linear Range: Where Instrument Response Is Reliable and Proportional
The analytical working range (also called linear range or measurement range) is the concentration/value interval over which an analytical instrument provides reliable, accurate, and proportional responses.
Key Characteristics
- Linearity: Signal output ∝ analyte concentration (typically R² ≥ 0.995).
- Precision: Repeatability (RSD) within validated limits (often ≤ 2–5%).
- Accuracy: Recovery within acceptance criteria (typically 95–105%).
- Lower Limit: Limit of Quantitation (LOQ) – lowest reliably quantifiable concentration.
- Upper Limit: Highest concentration before detector saturation or non-linearity.
Why It Matters
Operating outside this range yields:
- Below LOQ: High uncertainty, false negatives.
- Above Upper Limit: Signal plateau, underestimation, potential detector damage.
Best Practice: Dilute/concentrate samples to fall within the validated range. Document range in SOPs and validation reports (ICH Q2, USP <1225>, EPA methods).